SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z540-1-1994, AND ANSI/NCSL Z540.3-2006

Technical Maintenance, Inc.
117 Jetplex Circle,
Suite C4
Madison, AL 35758
Scott Chamberlain Phone: 256-772-4115

CALIBRATION

Chemical

Valid to: September 20, 2018 Certificate Number: AC-2080.02

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

pH2

(4, 7, & 10) pH units

0.019 pH

pH buffer solutions

Conductivity2

≈100 µS

≈1 410 µS

≈10 000 µS

0.069 µS

5.1 µS 34 µS

Conductivity solutions

Dimensional

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Calipers2,3

Up to 46 in

(26 + 10L) μin

Gage blocks (Grade 2)

Micrometers2,4

Up to 46 in

(30 + 6L) μin

Height Gages2,3

Up to 46 in

(200 + 3L) μin

Dial Indicators2,3

Up to 10 in

(5 + 57L) μin

Rulers2

Up to 46 in

0.0091 in

Metal Tapes and Rules2,3

Up to 100 ft

(0.000023L + 0.023) in

Standard rule

Feeler Gages2

Up to 1 in

73 μin

Mitutoyo 293-369 micrometer

Cylindrical gages – Plain Pin, Plugs, Rings2,3

Plain Pins, Plugs Rings

Up to 1 in (1 to 10) in

Up to 14 in

11 μin

(7 + 4D) μin

(8.0 + 2D) μin

P&W LabMaster gage blocks (grade 1)

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Surface Plates2

Overall Flatness Local Area Flatness

(18 × 18) in to

(6 × 6) ft

Up to (18 × 18) in

95 μin

74 μin

Rahn Planekator Repeat-o-meter

Gage Blocks

Up to 10 in

(2.9 + 2.2L) µin

Universal measuring

machine, master gage block set

Thread Plugs –

Major Pitch Diameter

Up to 12 in Up to 12 in

40 μin

91 μin

Gage blocks, P&W universal measuring machine,

VanKeren thread wire set

Optical Comparators2

Linearity

Magnification

Up to 20 in (20 to 40) in

10x to 100x

590 μin

790 μin

670 μin

Precision balls, Starrett Webber 81, SI Industries glass scales

Protractors2

(0 to 360) °

0.019 °

Angle blocks

Coating Thickness Gages2

Eddy Current & Magnetic Induction, Fixed Point

Coating Thickness Shims2

(0.48 to 38.9) mils

(0 to 243) mils

89 μin

80 μin

Calibration foils, P&W Supermicrometer

P&W Supermicrometer

Electrical – DC/Low Frequency

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

DC Voltage – Generate2

0 to 220 mV

(0.22 to 2.2) V

(2.2 to 11) V

(11 to 22) V

(22 to 220) V

(220 to 1 100) V

11 µV/V + 0.4 µV

5.2 µV/V + 0.7 µV

3.5 µV/V + 2.5 µV

3.5 µV/V + 4 µV

5 µV/V + 40 µV

6.5 µV/V + 0.4 mV

Fluke 5730A

DC Voltage – Measure2

Up to 100 mV (0.1 to 1) V

(1 to 10) V

(10 to 100) V

(100 to 1 000) V

11 μV/V + 0.3 μV

10 μV/V + 0.3 μV

10 μV/V + 0.5 μV

12 μV/V + 30 μV

12 μV/V + 0.1 mV

HP 3458A

(1 to 30) kV

(30 to 150) kV

0.13 %

0.12 %

Ross voltage dividers

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

DC Current – Generate2

Up to 220 μA (0.22 to 2.2) mA

(2.2 to 22) mA

(22 to 220) mA

(0.22 to 2.2) A

40 µA/A + 6 nA

35 µA/A + 7 nA

36 µA/A + 40 nA

48 µA/A + 0.7 µA

81 µA/A + 12 µA

Fluke 5730A

(2.2 to 11) A

(11 to 20.5) A

0.059 % + 0.5 mA 0.1 % + 0.75 mA

Fluke 5522A

DC Current – Measure2

Up to 100 nA (0.1 to 100) μA

100 μA to 10 mA

(10 to 100) mA

(0.1 to 1) A

31 μA/A + 0.04 nA

  1. μA/A + 0.8 nA

  2. μA/A + 50 nA

37 μA/A + 0.5 μA

0.011 % + 10 μA

HP 3458A

(1 to 600) A

0.3 %

Current Shunts

Electrical Calibration of Thermocouple Indicating Devices2

Type C Type J Type K Type N Type R Type S Type T

Type U

(0 to 2 316) °C

(-210 to 1 200) °C

(-200 to 1 372) °C

(-200 to 1 300) °C

(0 to 1 767) °C

(-250 to 400) °C

(-250 to 400) °C

(-200 to 600) °C

0.38 °C

0.19 °C

0.21 °C

0.27 °C

0.45 °C

0.44 °C

    1. °C

    2. °C

Fluke 7526A

Power Meters2

3 μW to 100 mW

0.25 %

HP 11683A

Resistance – Generate, Fixed Points2

1, 1.9

10, 19

100, 190

1, 1.9 k

10, 19 k

100 k

190 k

1 M

1.9 M 10 M 19 M

100 M

0.14 mΩ/Ω 33μΩ/Ω 35 μΩ/Ω 15 μΩ/Ω 13 μΩ/Ω 13 μΩ/Ω 15 μΩ/Ω 18 μΩ/Ω 25 μΩ/Ω 54 μΩ/Ω 66 μΩ/Ω

0.16 Ω/Ω

Fluke 5730A

Resistance – Generate2

Up to 11

(11 to 33)

(33 to 110)

(110 to 330)

330 to 1.1 k

(1.1 to 3.3) k

(3.3 to 11) k

(11 to 33) k

(33 to 110) kΩ

40 μΩ/Ω + 10 mΩ

30 μΩ/Ω + 15 mΩ

28 μΩ/Ω + 15 mΩ

28 μΩ/Ω + 0.02 Ω

29 μΩ/Ω + 0.02 Ω

29 μΩ/Ω + 0.2 Ω

29 μΩ/Ω + 0.1 Ω

29 μΩ/Ω + 1 Ω

29 μΩ/Ω + 1 Ω

Fluke 5520A

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Resistance2 – Measure

Up to 10

(10 to 100)

(0.1 to 1) k

(1 to 10) k

(10 to 100) k

(0.1 to 1) M

(1 to 10) M

(10 to 100) M

(0.1 to 1) G

16 μΩ/Ω + 50 μΩ

14 μΩ/Ω + 0.5 mΩ

12 μΩ/Ω + 0.5 mΩ

12 μΩ/Ω + 5 mΩ

12 μΩ/Ω + 5 mΩ

16 μΩ/Ω + 2.0 Ω

50 μΩ/Ω + 0.1 kΩ

0.05 % + 1 kΩ

0.51 % + 10 kΩ

HP 3458A

AC Voltage – Generate2

(0.22 to 2.2) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(2.2 to 22) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(22 to 220) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(0.22 to 2.2) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(2.2 to 22) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

0.36 mV/V + 5 µV

0.12 mV/V + 4 µV

0.11 mV/V + 4 µV

0.24 mV/V + 4 µV

0.6 mV/V + 5 µV

1.3 mV/V + 10 µV

1.7 mV/V + 20 µV

3.2 mV/V + 20 µV

0.36 mV/V + 5 µV

0.12 mV/V + 4 µV

0.11 mV/V + 4 µV

0.24 mV/V + 4 µV

0.6 mV/V + 5 µV

1.3 mV/V + 10 µV

1.7 mV/V + 20 µV

3.2 mV/V + 20 µV

0.24 mV/V + 12 µV

0.11 mV/V + 7 µV 69 µV/V + 7 µV

0.14 mV/V + 7 µV

0.37 mV/V + 17 µV

0.77 mV/V + 20 µV

1.7 mV /V + 25 µV

3.2 mV /V + 45 µV

0.28 mV/V + 40 µV

0.11 mV/V + 15 µV 50 µV/V + 8 µV 78 µV/V + 10 µV

0.1 mV/V + 30 µV

0.4 mV/V + 80 µV

1.2 mV /V + 0.2 mV

0.2 mV/V + 0.3 mV

0.28 mV/V + 0.4 mV

0.11 mV/V + 0.15 mV 50 µV/V + 50 µV 78 µV/V + 0.1 mV 98 µV/V + 0.2 mV

0.3 mV/V + 0.6 mV

1.2 mV/V + 2 mV

1.8 mV/V + 3.2 mV

Fluke 5730A

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

AC Voltage – Generate2

(22 to 220) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

0.28 mV/V + 4 mV

0.11 mV/V + 1.5 mV 62 µV/V + 0.6 mV 93 µV/V + 1 mV

0.18 mV/V + 2.5 mV

1.1 mV/V + 16 mV

5.1 mV/V + 40 mV

9.3 mV/V + 80 mV

Fluke 5730A

AC Voltage – Generate2

(220 to 750) V

(30 to 50_) KHz (50 to 100) kHz

(750 to 1 000) V

40 Hz to 1 KHz

(1 to 20) kHz

(20 to 30) KHz

0.7 mV/V + 11 mV

2.7 mV/V + 45 mV

0.11 mV/V + 4 mV

0.2 mV/V + 6 mV

0.7 mV/V + 11 mV

Fluke 5730A/5725A

AC Voltage – Measure2

Up to 10 mV

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

10 mV to 100 mV

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

(1 to 2) MHz

100 mV to 1 V

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

(1 to 2) MHz

1 V to 10 V

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

(1 to 2) MHz

0.044 % + 0.003 mV

0.026 % + 0.0011 mV

0.044 % + 0.0011 mV

0.11 % + 0.0011 mV

0.5 % + 0.0011 mV

4 % + 0.002 mV

0.019 % + 0.004 mV

0.019 % + 0.002 mV

0.027 % + 0.002 mV

0.045 % + 0.002 mV

0.09 % + 0.002 mV

0.31 % + 0.01 mV

1 % + 0.01 mV

1.5 % + 0.01 mV

0.019 % + 0.04 mV

0.019 % + 0.02 mV

0.027 % + 0.02 mV

0.045 % + 0.02 mV

0.09 % + 0.02 mV

0.31 % + 0.1 mV

1 % + 0.1 mV

1.5 % + 0.1 mV

0.019 % + 0.0004 V

0.019 % + 0.0002 V

0.027 % + 0.0002 V

0.045 % + 0.0002 V

0.09 % + 0.0002 V

0.31 % + 0.001 V

1 % + 0.001 V

1.5 % + 0.001 V

HP 3458A

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

AC Voltage – Measure2

(10 to 100) V

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

(100 to 700) V

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

0.026 % + 0.002 V

0.041 % + 0.002 V

0.038 % + 0.002 V

0.048 % + 0.002 V

0.13 % + 0.002 V

0.4 % + 0.01 V

1.5 % + 0.01 V

0.05 % + 0.04 V

0.05 % + 0.02 V

0.07 % + 0.02 V

0.13 % + 0.02 V

0.3 % + 0.02 V

HP 3458A

(1 to 21) kV

Up to 400 Hz (21 to 100) kV

Up to 400 Hz

0.59 %

0.62 %

Ross voltage dividers

AC Current – Generate2

(9 to 220) μA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(0.22 to 2.2) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(2.2 to 22) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(22 to 220) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(0.22 to 2.2) A

20 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

0.31 mA/A + 16 nA

0.21 mA/A + 10 nA

0.15 mA/A + 8 nA

0.35 mA/A + 12 nA

1.3 mA/A + 65 nA

0.3 mA/A + 40 nA

0.16 mA/A + 35 nA

0.14 mA/A + 35 nA 0.25 mA/A + 0.11 µA 1.3 mA/A + 0.65 µA

0.31 mA/A + 0.4 µA 0.2 mA/A + 0.35 µA 0.14 mA/A + 0.35 µA 0.26 mA/A + 0.55 µA

1.4 mA/A + 5 µA

0.3 mA/A + 4 µA 0.2 mA/A + 3.5 µA

0.14 mA/A + 2.5 µA

0.26 mA/A + 3.5 µA

1.4 mA/A + 10 µA

0.3 mA/A + 35 µA

0.54 mA/A + 80 µA

8.2 mA/A + 0.16 mA

Fluke 5730A

(2.2 to 11) A

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

560 µA/A + 0.17 mA

1200 µA/A + 0.38 mA

4300 µA/A + 0.75 mA

Fluke 5730A/5725A

(11 to 20.5) A

45 Hz to 100 Hz

100 Hz to 1 kHz

(1 to 5) kHz

0.19 % + 5 mA

0.24 % + 5 mA

4.6 % + 5 mA

Fluke 5522A

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

AC Current – Measure2

Up to 100 μA

(10 to 20) Hz

(20 to 45) Hz

45 Hz to 1 kHz

(1 to 100) mA

(10 to 20) Hz

(20 to 45) Hz

(45 to 100) Hz

100 Hz to 5 kHz

(5 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(0.1 to 1) A

(10 to 20) Hz

(20 to 45) Hz

(45 to 100) Hz

100 Hz to 5 Hz

(5 to 20) kHz

(20 to 50) kHz

0.41 % + 0.03 pA

0.16 % + 0.03 pA

0.07 % + 0.03 pA

0.41 % + 20 μA

0.16 % + 20 μA

0.069 % + 20 μA

0.038 % + 20 μA

0.069 % + 20 μA

0.41 % + 40 μA 0.56 % + 0.15 mA

0.41 % + 0.2 mA

0.17 % + 0.2 mA

0.087 % + 0.2 mA 0.11 % + 0.2 mA 0.31 % + 0.2 mA

1.0 % + 0.4 mA

HP 3458A

(1 to 600) A

30 Hz to 10 kHz

3.5%

Current Clamp

Capacitance2 – Generate

10 Hz to 10 kHz

10 Hz to 3 kHz

10 Hz to 1 kHz

(10 to 600) Hz

(10 to 300) Hz

(10 to 150) Hz

(10 to 120) Hz

(10 to 80) Hz

(0 to 50) Hz

(0 to 20) Hz

(0 to 6) Hz

(0 to 2) Hz

(0 to 0.6) Hz

(0 to 0.2) Hz

(0.19 to 0.1099) nF

(1.1 to 3.2999) nF

(3.3 to 329.999) nF

(0.33 to 1.09999) µF

(1.1 to 3.29999) µF

(3.3 to 10.9999) µF

(11 to 32.999) µF

(33 to 109.999) µF

(110 to 329.999) µF

(0.33 to 1.09999) mF

(1.1 to 3.2999) mF

(3.3 to 10.9999) mF

(11 to 32.9999) mF

(33 to 110) mF

1.2 % + 0.01 nF

1.2 % + 0.01 nF

0.39 % + 0.3 nF

0.38 % + 3 nF

0.37 % + 3 nF

0.38 % + 10 nF

0.53 % + 30 nF 0.58 % + 0.1 µF 0.58 % + 0.3 µF

0.57 % + 1 μF

0.58 % + 3 μF

0.59 % + 10 μF

0.65 % + 30 μF 1.3 % + 0.1 mF

Fluke 5522A

Oscilloscopes2,3

Amplitude – Square Wave (peak to peak)

50

1 M

Rise Time Flatness

(50 kHz reference)

5 mV to 5.5 V

Time Marker

1 mV to 6.6 V

1 mV to 130 V

<350 ps

50 kHz to 100 MHz

(100 to 300) MHz

(300 to 600) MHz

(600 to 1 100) MHz

1 ns to 20 ms

50 ms to 5 s

1.3 % + 40 µV

1.3 % + 40 µV

+ 13 ps / -0.11 ps

2.5 % + 0.1 mV

3.4 % + 0.1 mV

5.5 % + 0.1 mV

6.9 % + 0.1 mV

2.7 parts in 106 s

(25 + 1 000t) parts in 106 s

Fluke 5522A /SC1100

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Low Frequency Power –

Generate2

(45 to 65) Hz

1 PF

DC

Up to 20 kW

0.25 %

0.21 %

Fluke 5522A

Electrical – RF/Microwave

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

RF Attenuation2 – Measure

(-10 to 0) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-20 to -11) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-30 to -21) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-40 to -31) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-50 to -41) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-60 to -51) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

0.025 dB

    1. dB

    2. dB

0.032 dB

0.047 dB

0.033 dB

0.047 dB

0.035 dB

0.051 dB

0.037 dB

0.062 dB

0.04 dB

HP 8902A w/ HP 11722A, HP 11792A

RF Attenuation3 – Measure

(-70 to -61) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-80 to -71) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-90 to -81) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-100 to -91) dB

(10 to 1 300) MHz

(1.3 to 26.5) GHz

(-110 to -101) dB

(10 to 1 300) MHz

(-120 to -111) dB

(10 to 1 300) MHz

0.084 dB

0.081 dB

0.086 dB

0.13 dB

0.093 dB

0.14 dB

0.1 dB

0.16 dB

0.12 dB

0.24 dB

HP 8902A w/ HP 11722A, HP 11792A

RF Power3 – Measure 1 mW

50 MHz

0.0032 mW

HP 478A

RF Power – Generate

10 MHz to 40 GHz

(-10 to 10) dBm

(-60 to -10) dBm

(-110 to -60) dBm

1.5 dBm

1.3 dBm

1.8 dBm

HP 83460B opt 01

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Amplitude Modulation2

150 kHz to 10 MHz

(10 to 1300) MHz

(1.3 to 26.5) GHz

(5 to 99) %Depth

2.5 % + 1 digit

1.6 % + 1 digit

2.1 % + 1 digit

HP 8902A

Frequency Modulation2

250 kHz to 10 MHz

10 MHz to 26.5 GHz

(< 400 kHz)

2.1 % + 1 digit

1.2 % + 1 digit

HP 8902A

Phase Modulation2

150 kHz to 10 MHz

10 MHz to 26.5 GHz

200 Hz to 20 kHz

5.3 % + 1 digit

4.6 % + 1 digit

HP 8902A

RF Power2

100 kHz to 2.6 GHz

(0.1 to 26.5) GHz

(0 to 30) dBm

2 %

3.4 %

HP 8902A/HP 11722A HP 8902A/HP 11792A

AM/FM Distortion2

400 Hz and 1 kHz

1.2 dB

HP 8903A

Mechanical

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Scales & Balances2,3

(1 to 200) g

(0.022 to 2 000) lb

0.29 mg + 0.6R

0.01 % + 0.6R

Class 1 weights

Class F weights

Pressure2

(-14 to 5) psi

(5 to 10 000) psi

(0 to 30) psia

(0 to 100) cmH2O

(0 to 20) inH2O

(20 to 150) in·H2O

0.071 psi

0.12 %

0.019 psia

0.071 cmH2O 0.0018 inH2O

0.11 in·H2O

Fluke 700PD6

Ametek R-100 Fluke 700GA5 Heise 710B Additel 681-DP20

Additel 681-DP150

Vacuum

(0 to 30) inHg

0.12 inHg

Fluke 700PD6

Torque Wrenches2

(5 to 1 000) lbfin

(25 to 250) lbfft

(250 to 2 000) lbfft

0.35 %

0.43 %

0.49 %

CDI torque system

Torque Analyzers

(5 to 80) ozfin

(5 to 600) lbfin

(50 to 2 000 lbfft

0.17 %

0.15 %

0.14 %

Weights and Wheel

Mass

(1 to 10) lb

0.048 g

Master balance

Force – Tension2

(10 to 200) mgf

(0.2 to 1) gf

(1 to 10) gf

(10 to 500) gf

(1 to 540) lbs

0.63 mgf

1.0 mgf

0.038 %

0.025 %

0.026 %

Class F weights

 

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Accelerometers – Acceleration

(7 to 200) Hz

(100 to 2500) Hz

(2.5 to 10) kHz

(0.01 to 10) g

1.5 %

1.2 %

2.5 %

Accelerometer Calibrator

Thermodynamic

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Temperature – Measuring Equipment2

(50 to 600) °C

0.11 °C

Hart 1502A w/5626 PRT and dry block

Temperature – Measure3

(-25 to 600) °C

0.026 °C

Hart 1502 with 5626 PRT

Relative Humidity – Measure2

(0 to 90) % RH

1.6 % RH

Vaisalla HM141/HMP46

IR Thermometry2

(20 to 100) °C

(100 to 300) °C

(300 to 500) °C

0.51 °C

0.61 °C

0.8 °C

Fluke 9132

Time and Frequency

Parameter/ Equipment

Range

Calibration and Measurement Capability

[Expressed as Uncertainty (±)]

Reference Standard or Equipment

Frequency – Measure2

10 MHz

1 part in 1011 Hz/Hz

GPS 58503A/B

10 Hz to 500 MHz

5 parts in 107 Hz/Hz

HP 5345A

500 MHz to 26.5 GHz

(26.5 to 40) GHz

1.7 parts in 109 Hz/Hz 1 parts in 107 Hz/Hz

HP 5343A

Notes:

  1. Calibration and Measurement Capabilities (Expanded Uncertainties) are based on approximately a 95% confidence interval, using a coverage of k=2.

  2. This laboratory offers field calibration services

  3. L = length in inches, D = diameter in inches, t = time in seconds, R = resolution of device under test.

  4. % = percent of reading unless indicated otherwise.

  5. This scope is formatted as part of a single document including Certificate of Accreditation No. AC-2080.02

image