Melbourne, FL Certification

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z540-1-1994, and ANSI/NCSL Z540.3-2006
Technical Maintenance, Inc.
3060 Venture Lane
Melbourne, FL 32934
Scott Chamberlain Phone: 321-242-0890

CALIBRATION

Valid to: September 20, 2018 Certificate Number: AC-2080.08

I. Chemical

Parameter/ Equipment

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

pH3

(4, 7, 10) pH

0.02 pH

pH buffer solutions

Conductivity, Liquid3

(2 to 10 000) µS

1 % of reading

Conductivity solutions

II. Dimensional

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Gage Blocks

Up to 13 in

(1.7+2.1L) µin

Master gage blocks, P&W

universal measuring machine

Micrometers3

Up to 20 in

(32 + 4L) µin

Gage blocks (grade 2)

Calipers3

Up to 20 in

(48 + 7.2L) µin

Gage blocks (grade 2)

Dial Indicators3

Up to 10 in

(86 + 46L) µin

Gage blocks (grade 2)

Height Gages3

Up to 20 in

(194 + 3L) µin

Gage blocks (grade 2)

Rulers3

Up to 20 in

0.009 in

Gage blocks (grade 2)

Feeler Gage3

Up to 1 in

74 µin

Pratt & Whitney

Supermicrometer C



Version 001 Issued: 09/30/2016 Page 1 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

II. Dimensional

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Cylindrical Gages – Plain Pin, Plugs, Rings

Internal Diameter

(0.04 to 14) in

Outside Diameter

(0 to 13) in

(8 + 2D) μin

(2 + 3D) μin

Master gage blocks, P&W universal measuring machine

Thread Plugs3

Major

Pitch Diameter

Up to 12 in

Up to 12 in

40 µin

91 µin

B & S 599-246-00, Van

Keuren thread wire set

Gage blocks, P & W

universal measuring machine, Van Keuren thread wire set

Thread Rings

Up to 12 in

57 µin

Thread setting plug gages

Surface Plates3

Overall Flatness Only

Up to 6 ft  6 ft

50 μin

Mahr Federal level system

Optical Comparators3

Linearity

Magnification

Up to 20 in

(20 to 40) in

(10 to 100) X

590 μin

790 μin

590 μin

Gage blocks, SI Industries glass scales

III. Electrical DC/Low Frequency

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

DC Current – Generate3

Up to 220 µA

220 µA to 2.2 mA (2.2 to 22) mA

(22 to 220) mA

220 mA to 2.2 A

(2.2 to 11) A (11 to 20.5) A

40 µA/A + 6 nA

35 µA/A + 7 nA

36 µA/A + 40 nA

48 µA/A + 0.7 µA

81 µA/A + 12 µA

0.36 mA/A + 0.48 mA

0.76 mA/A + 0.75 mA

Fluke 5730A Fluke 5730A/5725A

Fluke 5522A

DC Current – Generate3

Clamp Only

(20.5 to 1 000) A

0.9 % of reading

Fluke 5522A/Coil

DC Current – Measure3

Up to 220 µA

220 µA to 2.2 mA

(2.2 to 22) mA (22 to 220) mA

(0.22 to 1) A

6 µA/A

5 µA/A

6 µA/A

10 µA/A

37 µA/A

Agilent 3458A



Version 001 Issued: 09/30/2016 Page 2 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

III. Electrical DC/Low Frequency

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

DC Voltage3 – Generate

(0 to 220) mV

220 mV to 2.2 V (2.2 to 11) V

(11 to 22) V (22 to 220) V

(220 to 1 100) V

11 µV/V + 0.4 µV

5.2 µV/V + 0.7 µV

3.5 µV/V + 2.5 µV

3.5 µV/V + 4 µV

5 µV/V + 40 µV

6.5 µV/V + 400 µV

Fluke 5730A

DC Voltage3 – Measure

Up to 100 mV

(0.1 to 1) V (1 to 10) V

(10 to 100) V (100 to 1 000) V

10 µV/V + 0.3 µV

9.0 µV/V + 0.30 µV

9.0 µV/V + 50 µV

11 µV/V + 30 µV

11 µV/V + 0.1 mV

Agilent 3458A/002

DC Voltage3 – Measure

(1 to 120) kV

0.23 % of reading

Ross VD120/ HP34401A

AC Voltage – Generate3

(0.22 to 2.2) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(2.2 to 22) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(22 to 220) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

220 mV to 2.2 V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

0.27 mV/V + 4 µV

0.14 mV/V + 4 µV

0.13 mV/V + 4 µV

0.22 mV/V + 4 µV

0.52 mV/V + 5 µV

1. 1 mV/V + 10 µV

1.4 mV/V + 20 µV

2.7 mV/V + 20 µV

0.24 mV/V + 4 µV

91 µV/V + 4 µV

81 µV/V + 4 µV

0.2 mV/V + 4 µV

0.5 mV/V + 5 µV

1.1 mV/V + 10 µV

1.4 mV/V + 20 µV

2.7 mV/V + 20 µV

0.24 mV/V + 12 µV

90 µV/V + 7 µV

58 µV/V + 7 µV

0.12 mV/V + 7 µV

0.31 mV/V + 17 µV

0.66 mV/V + 20 µV

1.4 mV /V + 25 µV

2.7 mV /V + 45 µV

0.24 mV/V + 40 µV

90 µV/V + 15 µV

43 µV/V + 8 µV

67 µV/V + 10 µV

85 µV/V + 30 µV

0.34 mV/V + 80 µV

1 mV /V + 0.2 mV

1.7 mV/V + 0.3 mV

Fluke 5730A/003



Version 001 Issued: 09/30/2016 Page 3 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

III. Electrical DC/Low Frequency

 

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

AC Voltage – Generate3

(2.2 to 22) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

0.24 mV/V + 0.4 mV

91 µV/V + 0.15 mV

43 µV/V + 50 µV

67 µV/V + 0.1 mV

83 µV/V + 0.2 mV

0.25 mV/V + 0.6 mV

1 mV/V + 2 mV

1 5 mV/V + 3.2 mV

AC Voltage – Generate3

(22 to 220) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(220 to 750) V

40 Hz to 1 KHz

(1 to 20) kHz

(20 to 50) KHz

(50 to 100) kHz

0.24 mV /V + 4 mV

90 µV/V + 1.5 mV

52 µV/V + 0.6 mV

80 µV/V + 1 mV

0.15 mV /V + 2.5 mV

91 µV/V + 4 mV

0.17 mV /V + 6 mV

0.6 mV /V + 11 mV

2 3 mV /V + 45 mV

Fluke 5730A/003

AC Voltage – Generate3

(750 to 1 000) V

40 Hz to 1 KHz

(1 to 20) kHz

(20 to 30) KHz

91 µV/V + 4 mV

0.17 mV /V + 6 mV

0.6 mV /V + 11 mV

Fluke 5730A/5725A

AC Voltage – Measure3

Up to 10 mV

(1 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz

10 mV to 10 V

(1 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

(1 to 2) MHz

(10 to 100) V

(1 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

0.03 % of reading + 0.03 % of range

0.02 % of reading + 0.011 % of range

0.03 % of reading + 0.011 % of range

0.1 % of reading + 0.011 % of range

0.5 % of reading + 0.011 % of range

4 % of reading + 0.02 % of range

0.007 % of reading + 0.004 % of range

0.007 % of reading + 0.002 % of range

0.014 % of reading + 0.002 % of range

0.03 % of reading + 0.002 % of range

0.8 % of reading + 0.002 % of range

0.3 % of reading + 0.01 % of range

1 % of reading + 0.01 % of range

1.5 % of reading + 0.01 % of range

0.02 % of reading + 0.004 % of range

0.02 % of reading + 0.002 % of range

0.02 % of reading + 0.002 % of range

0.04 % of reading + 0.002 % of range

0.12 % of reading + 0.002 % of range

0.4 % of reading + 0.002 % of range

1.5 % of reading + 0.01 % of range

Agilent 3458A/002



Version 001 Issued: 09/30/2016 Page 4 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

III. Electrical DC/Low Frequency

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

AC Voltage – Measure3

(100 to 1 000) V

(1 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz

(50 to 100) kHz

0.05 % of reading + 0.004 % of range

0.04 % of reading + 0.002 % of range

0.06 % of reading + 0.002 % of range

0.12 % of reading + 0.002 % of range

0.3 % of reading + 0.002 % of range

Agilent 3458A/002

AC Voltage – Measure3

(1 to 85) kV

(50, 60) Hz

0.7 % of reading

Ross VD120/ HP 34401A

AC Current – Generate3

(9 to 220) µA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(0.22 to 2.2) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(2.2 to 22) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

1 to 5 kHz

(5 to 10) kHz

(22 to 220) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

0.25 mA/A + 16 nA

0.16 mA /A + 10 nA

0.1 mA /A + 8 nA

0.28 mA /A + 12 nA

1.1 mA /A + 65 nA

0.25 mA /A + 40 nA

0.16 mA /A + 35 nA

0.1 mA /A + 35 nA

0.2 mA /A + 0.110 µA

1.1 mA /A + 0.650 µA

0.25 mA /A + 0.4 µA

0.16 mA /A + 0.35 µA

0.1 mA /A + 0.35 µA

0.2 mA /A + 0.55 µA

1.1 mA /A + 5 µA

0.25 mA /A + 4 µA

0.16 mA /A + 3.5 µA

0.1 mA /A + 2.5 µA

0.2 mA /A + 3.5 µA

1.1 mA /A + 10 µA

Fluke 5730A

AC Current – Generate3

(0.22 A to 2.2) A

20 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

0.25 mA/A + 35 µA

0.45 mA /A + 80 µA

7 mA /A + 0.16 mA

Fluke 5730A/5725A

AC Current – Generate3

(2.2 to 11) A

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(11 to 20.5) A

(45 to 100) Hz

100 Hz to 1 kHz

(1 to 5) kHz

0.46 mA /A + 0.17 mA

0.95 mA /A + 0.38 mA

3.6 mA /A + 0.75 mA

0.1 % of reading + 2 mA

0.12 % of reading + 5 mA

2.3 % of reading + 5 mA

Fluke 5522A

AC Current – Generate3

Clamp Only

(20.5 to 500) A

(45 to 440) Hz

1.8% of reading

Fluke 5522A/coil



Version 001 Issued: 09/30/2016 Page 5 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

III. Electrical DC/Low Frequency

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

AC Current – Measure3

Up to 100 µA

(10 to 20) Hz

(20 to 45) Hz

45 Hz to 1 kHz

(0.1 to 100) mA

(10 to 20) Hz (20 to 45) Hz (45 to 100) Hz

100 Hz to 5 kHz (5 to 20) kHz (20 to 50) kHz (50 to 100) kHz

(0.1 to 1) A

(10 to 20) Hz (20 to 45) Hz (45 to 100) Hz

100 Hz to 5 kHz (5 to 20) kHz (20 to 50) kHz

0.41 % of reading + 0.03 pA

0.16 % of reading + 0.03 pA

0.07 % of reading + 0.03 pA

0.41 % of reading + 20 μA

0.16 % of reading + 20 μA

0.07 % of reading + 20 μA

0.04 % of reading + 20 μA

0.08 % of reading + 20 µA

0.41 % of reading + 40 µA

0.56 % of reading + 150 µA

0.41 % of reading + 0.2 mA

0.17 % of reading + 0.2 mA

0.09 % of reading + 0.2 mA

0.11 % of reading + 0.2 mA

0.31 % of reading + 0.2 mA

1 % of reading + 0.4 mA

Agilent 3458A

Resistance – Generate, Fixed

Points3

1, 1.9 

10, 19 

100, 190 

1, 1.9 k

10, 19 k

100, 190 k

1 M

1.9 M

10 M

19 M

100 M

98 μΩ/Ω

23μΩ/Ω

10 μΩ/Ω

6.5 μΩ/Ω

6.5 μΩ/Ω

8.5 μΩ/Ω

13 μΩ/Ω

18 μΩ/Ω

40 μΩ/Ω

48 μΩ/Ω

0.1 mΩ/Ω

Fluke 5730A

Resistance – Generate3

Up to 11  (11 to 33)  (33 to 110) 

(110 to 330) 

330  to 1.1 k (1.1 to 3.3) k (3.3 to 11) k (11 to 33) k (33 to 110) kΩ

(110 to 330) kΩ (0.33 to 1.1) MΩ (1.1 to 3.3) MΩ (3.3 to 11) MΩ (11 to 33) MΩ (33 to 110) MΩ (110 to 330) MΩ

330 MΩ to 1.1 GΩ

40 μΩ/Ω + 0.01 Ω

30 μΩ/Ω + 0.015 Ω

28 μΩ/Ω + 0.015 Ω

28 μΩ/Ω + 0.02 Ω

29 μΩ/Ω + 0.02 Ω

29 μΩ/Ω + 0.2 Ω

29 μΩ/Ω + 0.1 Ω

29 μΩ/Ω + 1 Ω

29 μΩ/Ω + 1 Ω

33 μΩ/Ω + 10 Ω

33 μΩ/Ω + 10 Ω

60 μΩ/Ω + 150 Ω

0.013 % of reading + 0.25 kΩ

0.025 % of reading + 2.5 kΩ

0.05 % of reading + 3.0 kΩ

0.3 % of reading + 100 kΩ

1.5 % of reading + 500 kΩ

Fluke 5522A



Version 001 Issued: 09/30/2016 Page 6 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

III. Electrical DC/Low Frequency

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Resistance – Measure3

Fixed Points

Up to 10 

10 to 100 

100 to 100 k

100 k to 1 M

1 M to 10 M

10 M to 100 M

100 M to 1 G

16 μΩ/Ω + 50 µΩ

14 μΩ/Ω + 0.5 mΩ

12 μΩ/Ω + 5 mΩ

16 μΩ/Ω + 2 Ω

50 μΩ/Ω + 100 Ω

0.05% of reading + 1.0 kΩ

9.2 μΩ/Ω + 1 Ω

Agilent 3458A

Capacitance – Generate3

(0.19 to 0.1099) nF

(1.1 to 3.299) nF (3.3 to 329.999) nF (0.33 to 1.09999) µF (1.1 to 3.29999) µF (3.3 to 10.9999) µF (11 to 32.999) µF

(33 to 109.999) µF (110 to 329.999) µF (0.33 to 1.09999) mF (1.1 to 3.2999) mF (3.3 to 10.9999) mF (11 to 32.9999) mF (33 to 110) mF

1.2 % of reading + 0.01 nF

1.2 % of reading + 0.01 nF

0.39 % of reading + 0.3 nF

0.38 % of reading + 3 nF

0.37 % of reading + 3 nF

0.38 % of reading + 10 nF

0.53 % of reading + 30 nF

0.58 % of reading + 0.1 µF

0.58 % of reading + 0.3 µF

0.57 % of reading + 1 µF

0.58 % of reading + 3 µF

0.59 % of reading + 10 µF

0.65 % of reading + 30 µF

1.3 % of reading + 0.1 mF

Fluke 5522A

Oscilloscope Calibration –

Vertical Deflection3

50 , 1 kHz square wave

1 M, 1 kHz square wave

Flatness – Leveled Sine Wave

5 mV to 5.5 V

Bandwidth

Rise Time

Time Interval

1 mV to 6.6 V

1 mV to 130 V

50 kHz to 100 MHz (100 to 300) MHz (300 to 600) MHz (600 to 1 100) MHz

<300 pS

1 nS to 20 mS

50 mS to 5 S

0.59 % of reading + 40 µV

0.44 % of reading + 40 µV

2.5 % of reading + 0.1 mV

2.2 % of reading + 0.1 mV

4.6 % of reading + 0.1 mV

6.4 % of reading + 0.1 mV

+ 13 pS / -110 pS

3 μS/S

(26 + 1000t) μs/S

Fluke 5522A/SC1100

Power Meter Range Calibration3

3 μW to 100 mW

0.25 % of reading

HP 11683A



Version 001 Issued: 09/30/2016 Page 7 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

III. Electrical DC/Low Frequency

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Electrical Calibration of Thermocouple Indicating Devices3

Type B Type C Type E Type J Type K Type L Type N Type R Type S Type T Type T

Type U

(600 to 1 820) °C (0 to 2 316) °C

(-250 to 1 000) °C (-210 to 1 200) °C (-200 to 1 372) °C (-200 to 900) °C

(-200 to 1 300) °C (0 to 1 767) °C

(0 to 1 767) °C

(-250 to 200) °C (-200 to 400) °C

(-200 to 600) °C

0.41 °C

0.38 °C

0.29 °C

0.19 °C

0.21 °C

0.12 °C

0.26 °C

0.45 °C

0.44 °C

0.41 °C

0.20 °C

0.19 °C

Fluke 7526A

IV. Electrical RF/Microwave

Parameter/ Equipment

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Flatness – Measure

Up to 7 V

30 MHz to 50 GHz

6 % of reading

8482A, N5532A

AM Distortion3 – Measure

(0.1 to 10) MHz

10 MHz to 26.5 GHz

0.8 % of reading

1 % of reading

Agilent N5531S

FM Distortion3 – Measure

1 MHz to 26.5 GHz

0.3 % of reading

Agilent N5531S

Phase Modulation3

Measure

100 kHz to 26.5 GHz

9.7 % of reading

Agilent N5531S



Version 001 Issued: 09/30/2016 Page 8 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

IV. Electrical RF/Microwave

Parameter/ Equipment

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Attenuation3 – Measure

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 26.5 GHz

10 MHz to 3.05 GHz

(0 to -10) dB

(-11 to -20) dB (-21 to -30) dB (-31 to -40) dB (-41 to -50) dB (-51 to -60) dB (-61 to -70) dB (-71 to -80) dB (-81 to -90) dB

(-91 to -100) dB

(-101 to -110) dB (-111 to -120) dB

0.06 dB

0.06 dB

0.06 dB

0.06 dB

0.06 dB

0.07 dB

0.06 dB

0.08 dB

0.08 dB

0.08 dB

0.08 dB

0.08 dB

Agilent N5531S

Amplitude

Modulation3 – Measure

100 kHz to 10 MHz

10 MHz to 3 GHz

10 MHz to 3 GHz

(3 to 26.5) GHz

(3 to 26.5) GHz

(5 to 99) % Depth (5 to 20) % Depth (20 to 99) % Depth (5 to 20) % Depth (20 to 99) % Depth

1 % of reading

2.9 % of reading

0.8 % of reading

5.2 % of reading

1.8 % of reading

Agilent N5531S

Frequency Modulation3

Measure

250 kHz to 10 MHz

10 MHz to 6.6 GHz (6.6 to 13.2) GHz (13.2 to 26.5) GHz

20 Hz to 10 kHz

(50 to 200) Hz

3.1 % of reading

3.1 % of reading

3.8 % of reading

5 % of reading

Agilent N5531S

RF Power – Generate

10 MHz to 40 GHz

10 MHz to 40 GHz

10 MHz to 40 GHz

(10 to -10) dBm

(-10 to -60) dBm

(-60 to -110) dBm

1.3 dBm

1.7 dBm

2.4 dBm

Keysight E8257D

RF Power3 – Measure

50 MHz

100kHz to 4.2GHz

10MHz to 18GHz

10MHz to 26.5GHz

1 mW

(-30 to 20) dBm

(-20 to 30) dBm

0.0032 mW

3.1 % of reading

3.1 % of reading

4.7 % of reading

HP 478A HP 8482A

HP 8481A/E9300A

Agilent N5531S

Phase Noise3 – SSB Measure

1 MHz to 50 GHz

0.39 dB

Agilent E4440A

AM/FM Distortion3

20 Hz to 20 kHz

1.2 dB

HP 8903B

Harmonic Distortion

20 Hz to 200 kHz

200 kHz to 26.5 GHz

1.2 dB

1.2 dB

HP 8903B

Agilent E4448



Version 001 Issued: 09/30/2016 Page 9 of 10
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org

V. Mechanical

Parameter/ Equipment2

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Scales & Balances3

(0.5 to 500) lb

0.013% of reading + 0.6R

Class F Weights

Force – Tension and

Compression3

(0.5 to 500) lbf

0.09 % of reading

Class F weights

Pressure3

(-15 to 100) psi

(100 to 1 000) psi

(1 000 to 15 000) psi

0.28 psi

0.26 psi

9 psi

Additel ADT681-20-GP100- PSI-N

Additel ADT681-02-GP1K- PSI-N

Additel ADT681-05-

GP15K-PSI-N

Torque Tools3

4 lbf⋅in to 600 lbf⋅ft

0.32 % of reading

CDI 5000 ST torque tester

VI. Thermodynamics

Parameter/ Equipment

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Relative Humidity –

Measure

(0 to 90) % RH

1.7 % RH

Vaisala MI70 / HMP75

Temperature –

Measuring Equipment3

(-25 to 350) °C

0.12 °C

Hart 1502A with 5616 PRT

and dry block

Temperature – Measure3

(-25 to 600) °C

0.025 °C

Hart 1502A with 5616 PRT

IR Thermometry3

(20 to 100) ºC

(100 to 300) ºC (300 to 500) ºC

0.51 ºC

0.61 ºC

0.80 ºC

Fluke 9132

&‌epsilon; = 0.95

VI. Time and Frequency

Parameter/ Equipment

Range

Calibration and Measurement

Capability

[Expressed as Uncertainty (±)]

Reference Standard or

Equipment

Frequency – Measure

10 MHz

1 x 10-11 Hz/Hz

58503A/B GPS

Frequency – Measure3

10 Hz to 26.5 GHz

1 x 10-8 Hz/Hz

HP 53131A, E4440A

Notes:

1. Calibration and Measurement Capabilities (Expanded Uncertainties) are based on approximately a 95% confidence interval, using a coverage of k=2.
2. L = length in inches, t = time in seconds, D = diameter in inches, R = resolution of the device under test.
3. On-site calibration service available for this measurement parameter.
4. This scope is formatted as part of a single document including Certificate of Accreditation No. AC-2080.08



Version 001 Issued: 09/30/2016 Page 10 of 10